中文

UV-int150 Standard UV Energy Meter

Industry Standard Ultraviolet Energy Measurement Instrument | Standard Tool in UV Curing Industry Since 2004

Product Overview

Industry Standard Product

Since its design in 2004, UV-int150 has become the standard measurement tool in UV curing industry, widely used in electronics manufacturing, printing, coating, medical equipment and other industries for quality control and process optimization.

Core Features

Product Introduction

The UV-int150 standard UV energy meter is a high-quality ultraviolet energy measurement instrument specifically designed for measuring UV energy of UV curing equipment. The instrument features a standard design, making it convenient for use in industrial production environments and providing accurate ultraviolet energy measurement data.

The main sensor is located on the back, with a measurement range of 250~410 nanometers. Measurement results are directly displayed on the 6-digit LCD screen on the front, in millijoules per square centimeter.

Passed authoritative metrology calibration report from NIM

UV-int150 Standard UV Energy Meter

UV-int150 Standard UV Energy Meter - Authentication: +86-13306512129

Recommended:UV Warning UV-int140

Technical Specifications

Basic Parameters

  • Spectral Range: 250-410 nanometers
  • Measuring Range: 0-10,000 mW/cm² (upgraded to 2x original)
  • Display Range: 6-digit LCD, 0-999,999 mJ/cm²
  • Dimensions: Diameter 90mm, Height 12mm
  • Weight: Approximately 150 grams

Performance Indicators

  • Measurement Accuracy: ±5%
  • Maximum Temperature: 110℃
  • Power Supply: 3.6V lithium battery
  • Power Consumption: 100μA, 10,000 hours usage
  • Housing: Aluminum casing

Application Features

  • Suitable for UV curing processes
  • Ideal for UV coatings and UV adhesives testing
  • Industrial-grade quality inspection
  • Process optimization and quality control
  • Scientific research and teaching demonstrations

Installation & Operation

Operation Steps

  1. Turn ON the switch, the LCD display shows zero
  2. Place the UV energy meter near the UV lamp source, try to align its back toward the light source (since the energy sensor is on the back), extend the exposure time appropriately
  3. Wait until the meter's display shows measurement results, for example 800 millijoules per square centimeter (800 mJ/cm²)
  4. Turn OFF the energy meter switch (OFF), or reset the display value to zero

Precautions

UV-int150 PDF Documentation Download

Contact Us

Learn more about the UV-int150 standard UV energy meter

Order Now View Product Catalog View High Temperature UV-int150+

Consultation Hotline: +86-13306512129 Mr. Hu